OUTPUT
Voltage/Current Ranges
2 x (0 to >75V; 0 to >2A) (300W nominal power)
Multi-Mode Operation
Three modes of operation can be selected via a
rotary switch:
Mode A: Two independent and isolated outputs
(75V/2A each).
Mode B: One output of double the current
capability (75V/4A).
(Unused output is disabled and its displays are blanked).
Mode C: One output of double the voltage
capability (150V/2A).
(Unused output is disabled and its displays are blanked).
Output Setting & Control
Voltage Setting: By coarse and fine controls.
Current Setting: By single logarithmic control.
Output Mode: Constant voltage or constant
current with automatic cross-over. CC indicator in
constant current mode.
Output Switch: Electronic, non isolating. Preset
voltage and current limit displayed when Output off.
Output Performance
Ripple & Noise: Typically <2mV rms, <15mV pk-pk
(CV mode, 20MHz bandwidth)
Load Regulation: Modes A and B - <0·01%; Mode C
- <0·1% + 2mV (90% load change)
Line Regulation: <0·01% of maximum output for
10% line change.
Transient Response: <200µs to within 50mV of
setting for a 5% to 95% load change.
Temp. Coefficient: Voltage: typically
<100ppm/°C.
Output Protection
Output Protection: Outputs will withstand
forward voltages up to 85V/170V.
Over-voltage trip operates above these levels. Reverse protection
by diode clamp for currents to 3A.
Output Connections
Output Terminals: High voltage touch-proof
terminals.
Terminals can accept standard 4mm plugs, fork
terminals and bare wires.
METER SPECIFICATIONS
Display Type: 3 digit meters for voltage and
current;
(4 digit on second voltmeter to maintain resolution above 100V).
Voltage Meter
Resolution: 100mV
Accuracy: 0·3% of reading ± 1 digits
Current Meter
Resolution: 10mA
Accuracy: 0·6% of reading ±
1 digits
GENERAL
Input
AC Input: 110V to 240V 50/60Hz (universal input)
VA Rating: 500VA
Temperature & Environmental
Operating Range: +5ºC to +40ºC, 20% to 80% RH
Storage Range: -40ºC to + 70ºC
Environmental: Indoor use at altitudes up to
2000m, Pollution Degree 2.
Cooling: fan-less convection cooling
Safety & EMC
Safety: Complies with EN61010-1
EMC: Complies with EN61326
Physical
Size: 260x160x320 mm (WxHxD).
Weight: 4.3kg. |